Decade boxes put a precise, switchable value at your fingertips: resistance decades for simulating sensors and checking meters, capacitance and inductance decades for filter and bridge work. They are often treated as permanent references, but switch contacts wear, resistors age and a single damaged decade can affect every setting that uses it.
Decade box calibration measures the box across its decades and switch positions so you know how far each setting is from its nominal value.
What gets measured
- Each decade at several switch positions, not just one setting per box. Because decades add in series, an error in one decade carries into every combined value.
- Zero (residual) value: the resistance, capacitance or inductance present with every dial at zero. It is often the largest correction at low settings.
- Switch repeatability: worn contacts give different values each time a dial is turned.
- Frequency (capacitance and inductance boxes): values are measured at the frequencies you use, because they change with frequency.
Two-wire versus four-wire
For low resistance settings, lead and contact resistance can be larger than the error you are trying to see. Precision resistance decades with separate sense terminals should be measured four-wire, and the certificate should say which connection was used. Two-wire results are only meaningful if you also use the box two-wire.
Using the results
If you use a decade box as a reference, apply the measured corrections rather than the nominal values, and include the box’s uncertainty in your own uncertainty budget. Keep the box at a stable temperature, because resistor values drift with temperature, and give it time to settle after transport.