Linear variable differential transformers (LVDTs) measure displacement in test rigs, gauging fixtures, servo valves and structural tests. They are robust and have no contacting parts in the sensing element, but the reading depends on the whole chain: the transducer, its signal conditioner and the way it is mounted.
LVDT calibration establishes the relationship between core position and output across the working stroke and documents how linear it is.
How an LVDT is calibrated
The LVDT body is clamped in a fixture, and the core is moved through known displacements using a micrometer-head calibrator, a precision positioning stage or gage blocks. At each point, the output from the conditioner is recorded. Measurements are taken across the full stroke and in both directions, so that hysteresis and backlash in the setup show up.
What the results tell you
- Sensitivity: output per unit of displacement, for example mV/V per mm, used to scale readings.
- Linearity: the largest deviation from a best-fit straight line, usually stated as a percentage of full stroke. Linearity worsens near the ends of the stroke, so the certificate should show where the useful range ends.
- Null position: the core position giving zero output, which matters for bidirectional measurements.
- Repeatability across repeated runs.
Calibrate the system, not just the sensor
LVDT output depends on excitation frequency and amplitude, cable length and the conditioner’s gain and filtering. Calibrating the transducer and its conditioner together, as used, gives a result you can apply directly. If you swap conditioners or cables, recheck at least the sensitivity. Mounting also matters: a side-loaded core or a misaligned body adds errors that no calibration will remove.
Intervals and related equipment
Annual calibration is a common starting point, with checks after any crash, cable damage or conditioner change. LVDTs are often used alongside dial and digital indicators and load cells on the same test rigs.
Accreditation and certificates
LVDT calibration is not listed on Techmaster’s ANAB scope, so it is calibrated with NIST-traceable reference standards and issued a traceable certificate with as-found and as-left data. If your auditor requires an ISO/IEC 17025 accredited certificate, tell us before you send the instrument.