A Class II biological safety cabinet protects three things at once: the operator, the product and the environment. It does that with a carefully balanced curtain of inflow air at the front, HEPA-filtered downflow over the work surface and filtered exhaust. If any one of those drifts, the cabinet can look and sound normal while it no longer protects anything.

biological safety cabinet certification — Techmaster ISO 17025 accredited calibration
Biological Safety Cabinet Certification at Techmaster Electronics — ANAB-accredited (AC-1736), NIST-traceable, ISO/IEC 17025, in-lab or on-site across the USA.

Biological safety cabinet certification checks each of those airflows and the filters against NSF/ANSI 49, and records the results for your biosafety program.

The field tests in NSF/ANSI 49

TestWhat it checks
Inflow velocityAir entering through the front opening, which keeps aerosols inside the cabinet and protects the operator
Downflow velocityThe filtered air moving down over the work surface, which protects the product; measured on a grid and checked for uniformity
HEPA filter leak testAn aerosol challenge upstream of the supply and exhaust filters, scanned with a photometer to find leaks in the media, seals or frame
Airflow smoke patternSmoke shows that air at the front opening moves into the cabinet and does not escape or reflux onto the work surface
Site installation checksExhaust connection, alarms and the cabinet’s position relative to doors and supply diffusers

When certification is required

The CDC/NIH guidance in Biosafety in Microbiological and Biomedical Laboratories (BMBL) calls for Class II cabinets to be certified when they are installed, at least once a year after that, and whenever they are moved or repaired. Moving a cabinet, even across the same room, can damage filter seals and change the airflow balance. Replacing a filter or motor always calls for recertification before use.

Cabinet class and type matter

Class II cabinets come in several types. A1 and A2 cabinets recirculate most of their air and exhaust the rest, while B1 and B2 cabinets are hard-ducted to the building exhaust. The type changes what the test has to confirm. For a B2 cabinet, for example, the building exhaust must be proven to carry the full exhaust load, and the cabinet alarm must respond if it fails. The certification report should name the cabinet’s class, type and the acceptance values used.

Preparing for certification

  • Decontaminate the cabinet if filters or internal panels will be accessed, and provide the decontamination record.
  • Remove stored equipment and materials from the work area.
  • Have the manufacturer’s nominal inflow and downflow values available, along with the previous report.
  • Plan for the cabinet to be out of use during testing.

Related containment testing

Cabinets are often certified on the same visit as fume hoods and cleanroom spaces. See fume hood testing and cleanroom testing and certification.

Accreditation and certificates

Biological safety cabinet field testing to NSF/ANSI 49 is on Techmaster’s ANAB ISO/IEC 17025 scope (certificate AC-1736), so certificates carry the ANAB accreditation symbol along with as-found and as-left data and measurement uncertainty.

Frequently asked questions

How often does a biological safety cabinet need certification?
At installation, at least annually after that, and whenever the cabinet is moved, repaired or has a filter replaced, in line with the CDC/NIH BMBL guidance.
What tests are part of BSC certification?
Inflow velocity, downflow velocity, HEPA filter leak testing, airflow smoke pattern tests and site installation and alarm checks, following NSF/ANSI 49.
What is the difference between Class II A2 and B2 cabinets?
A2 cabinets recirculate most of their air through HEPA filters and exhaust the rest, while B2 cabinets are hard-ducted and exhaust all of their air to the building.
Is Techmaster's BSC certification accredited?
Yes. Biological safety cabinet testing to NSF/ANSI 49 is on Techmaster's ANAB ISO/IEC 17025 scope (certificate AC-1736).

References & industry standards

External standards bodies. Techmaster Electronics is an ISO/IEC 17025-accredited, NIST-traceable calibration laboratory.

Accredited calibration at Techmaster

Techmaster Electronics is an ANAB-accredited (Certificate AC-1736) ISO/IEC 17025:2017 calibration laboratory with five US labs, NIST-traceable standards and a full measurement-uncertainty budget on every certificate — in-lab or nationwide on-site, with 5-business-day standard turnaround and expedite options. See our calibration services or request a quote.

Khanh Nguyen

Khanh Nguyen

Khanh Nguyen is the Marketing Manager at Techmaster Electronics, a B2B marketing leader covering the test, measurement, and ISO/IEC 17025 accredited calibration industry in the United States.

Request a Quote