A quality manager is setting up daily checks on a new 220 g analytical balance and a 30 kg bench scale. The catalog of calibration weights offers the same 100 g weight in eight grades at eight prices, and a 20 kg cast iron weight marked “Class F” for a fraction of the cost of the stainless one. Which is good enough, and which will an auditor question?
The answer depends on the test weight class, the tolerance you hold the weighing device to, and whether you use the weight at its nominal value or at the value on its certificate. This guide puts the three weight systems used in the United States side by side with numbers from the current standards, then works through the selection rule with two examples. Techmaster Electronics is an ISO/IEC 17025 calibration laboratory founded in 1989 and accredited by ANAB under Certificate AC-1736, with accredited laboratories in Vista and Santa Clara, California, Orlando, Florida, and San Antonio, Texas.
What does a test weight class actually tell you?
The class mark covers three things:
- The tolerance applies to conventional mass. OIML R 111 defines conventional mass at 20 °C, against a reference density of 8,000 kg/m³, in air of 1.2 kg/m³. That convention is why a stainless steel weight and a cast iron weight of the same class can be used on the same scale. For the cases where it breaks down, see our guide to air buoyancy correction in mass calibration at altitude.
- The class limits the certificate uncertainty too. Both ASTM E617 and OIML R 111 require the expanded uncertainty of a weight’s calibration to be no more than one-third of its class tolerance. A weight cannot be declared Class 1 by a laboratory whose uncertainty is as large as the Class 1 tolerance.
- The class sets how the weight is built. In ASTM E617-23, Classes 000, 00 and 0 must be one-piece (Type I) weights. Classes 1 to 7 may contain adjusting material behind a plug or knob (Type II). Cast iron is confined to the coarse classes and larger sizes.
A class is not a calibration. A weight sold as Class 2 with no certificate has a claimed tolerance and no evidence. A weight with an accredited certificate has a measured value, an uncertainty and a date, and that is what your traceability rests on.
How do ASTM E617, OIML R 111 and NIST Class F compare?
The table below lists the maximum tolerance for three common nominal values. ASTM figures are from E617-23, OIML figures are from OIML R 111-1 (2004), Table 1, and Class F figures are from the 1990 edition of NIST Handbook 105-1.
| Class | System | 100 g | 1 kg | 20 kg |
|---|---|---|---|---|
| ASTM 000 | ASTM E617-23 | 0.025 mg | 0.25 mg | 5.0 mg |
| ASTM 00 | ASTM E617-23 | 0.05 mg | 0.5 mg | 10 mg |
| OIML E1 | OIML R 111 | 0.05 mg | 0.5 mg | 10 mg |
| ASTM 0 | ASTM E617-23 | 0.13 mg | 1.3 mg | 25 mg |
| OIML E2 | OIML R 111 | 0.16 mg | 1.6 mg | 30 mg |
| ASTM 1 | ASTM E617-23 | 0.25 mg | 2.5 mg | 50 mg |
| ASTM 2 | ASTM E617-23 | 0.50 mg | 5.0 mg | 100 mg |
| OIML F1 | OIML R 111 | 0.5 mg | 5.0 mg | 100 mg |
| ASTM 3 | ASTM E617-23 | 1.0 mg | 10 mg | 200 mg |
| OIML F2 | OIML R 111 | 1.6 mg | 16 mg | 300 mg |
| ASTM 4 | ASTM E617-23 | 2.0 mg | 20 mg | 400 mg |
| OIML M1 | OIML R 111 | 5.0 mg | 50 mg | 1.0 g |
| ASTM 5 | ASTM E617-23 | 10 mg | 50 mg | 1.0 g |
| ASTM 6 | ASTM E617-23 | 20 mg | 100 mg | 2.0 g |
| NIST Class F | Handbook 105-1 (1990) | 20 mg | 100 mg | 2.0 g |
| OIML M2 | OIML R 111 | 16 mg | 160 mg | 3.0 g |
| ASTM 7 | ASTM E617-23 | 100 mg | 470 mg | 3.8 g |
| OIML M3 | OIML R 111 | 50 mg | 500 mg | 10 g |
Tolerances are plus or minus. Rows are ordered by the 1 kg tolerance. Check the edition before you copy a number: see the note on Classes 5 and 6 below.

There is no official ASTM-to-OIML equivalence. ASTM Class 00 and OIML E1 match at these three nominal values, and ASTM Class 2 matches OIML F1, but the match does not hold at every size. ASTM Class 1, the class most often bought for analytical balances, sits between OIML E2 and F1. If a procedure calls for “F1 or better,” a Class 1 weight meets it at these sizes. If it calls for E2, Class 1 does not.
The order changes with size. At 100 g, OIML M1 (5 mg) is tighter than ASTM Class 5 (10 mg). At 1 kg and 20 kg they are equal. Always compare at the nominal value you will use.
ASTM E617-23 changed Classes 5 and 6. Many tolerance charts still in circulation show the 2018 values. For a 100 g weight, Class 6 went from 10 mg to 20 mg, and for a 1 g weight it went from 2.0 mg to 0.90 mg. The current edition is ASTM E617-23. If your procedure cites Class 5 or 6 limits, cite the edition next to them.
How tight does a test weight need to be for your scale?
The rule appears in several places that a US quality system is likely to cite:
- NIST Handbook 44. Appendix A, section 3.2 of the current edition of NIST Handbook 44 says the combined error and uncertainty of a standard used without correction “must be less than one-third of the applicable device tolerance.” If a correction is applied, the uncertainty of the corrected value has to meet the same limit.
- USP General Chapter 41. The chapter limits balance accuracy to 0.10%. USP’s FAQ allows the weight’s nominal value to be used when its maximum permissible error is under one-third of the limit applied. The chapter was revised with an official date in February 2026, so check the current USP-NF for the limit that applies to each test. Our USP 41 minimum weight guide for analytical balances covers the repeatability side.
- The weight standards themselves. ASTM E617 and OIML R 111 apply the same one-third ratio one level up, between a weight’s tolerance and the uncertainty of its calibration.
Example A: an analytical balance checked at 100 g
Your procedure holds the balance to ±0.5 mg at 100 g. One-third of that is 0.17 mg. On tolerance alone, ASTM Class 0 (0.13 mg) and OIML E2 (0.16 mg) fit, and ASTM Class 1 (0.25 mg) and OIML F1 (0.5 mg) do not. Under the stricter Handbook 44 wording, which adds the weight’s calibration uncertainty to its error, Class 0 and E2 are marginal here, so use the calibrated value or step up a class.
Now change the tolerance. If the only requirement is an accuracy limit of 0.10%, the limit at 100 g is 100 mg and one-third is 33 mg. An ASTM Class 4 weight (2.0 mg) is more than adequate. USP’s FAQ makes the same point with a 200 g weight checked against a 0.05% limit: an OIML F2 weight is enough and an E2 or ASTM Class 1 weight is not needed. The same balance needs a very different weight depending on the tolerance written in the procedure.
Example B: a 30 kg bench scale checked at 20 kg
The scale reads in 1 g steps and your procedure allows ±2 g at 20 kg. One-third is 0.67 g. ASTM Class 4 (0.40 g) fits. OIML M1 and ASTM Class 5 (1.0 g each) do not, and a cast iron ASTM Class 6 or NIST Class F weight (2.0 g) is as loose as the tolerance you are trying to check.

Cast iron Class F weights are sold as “scale test weights,” and they are appropriate for coarser scales, where the tolerance at 20 kg is above about 8 g. On a 30 kg scale that resolves 1 g, they are not good enough at nominal value. That leads to the next question.
Can you use a lower class if you use the calibrated value?
EURAMET Calibration Guide No. 18, a widely used reference for balance calibration, treats the two cases differently. When a weight is used at nominal value within a tolerance, its standard uncertainty is taken as the tolerance divided by the square root of three. When its calibrated value is used, the standard uncertainty is the certificate’s expanded uncertainty divided by the coverage factor, with a separate term for drift since the last calibration.
Take a 1 kg ASTM Class 1 weight. Used at nominal value, its 2.5 mg tolerance contributes about 1.4 mg of standard uncertainty. E617 caps the certificate uncertainty for that weight at one-third of the tolerance, 0.83 mg at k = 2, so used at its calibrated value it contributes no more than about 0.42 mg before drift. That is a factor of three or more from the same weight.
That approach has conditions:
- The operator has to enter the certificate value for that serial number, every time.
- The value expires with the certificate. A dropped or corroded weight no longer matches it.
- Drift between calibrations becomes part of your ISO/IEC 17025 measurement uncertainty budget, and you need more than one calibration to estimate it.
For a daily user check on a production floor, a weight that passes the one-third rule at nominal value is the more reliable choice. For a laboratory calibrating balances, using calibrated values is normal practice.
Which test weight class fits which scale or balance?
| Weighing device | Typical readability | Common ASTM E617 class | Common OIML R 111 class |
|---|---|---|---|
| Microbalance, semi-micro balance | 0.001 mg to 0.01 mg | 000, 00, 0 | E1, E2 |
| Analytical balance (Handbook 44 Class I) | 0.1 mg | 1 | E2, F1 |
| Precision top-loading balance (Class II) | 0.001 g to 0.01 g | 1, 2 | F1, F2 |
| Moderate-precision balance | 0.01 g to 0.1 g | 3 | F2, M1 |
| Bench, counting and floor scales (Class III, IIIL) | 1 g and coarser | 3, 4, 5, 6 | M1, M2 |
ASTM and OIML classes for Handbook 44 Classes I to IIIL follow Table 1 of NIST Handbook 105-1 (2019). The microbalance row and the E2 entry come from weight manufacturers’ selection guidance. These are common pairings, not requirements.
The table gets you to the right shelf. The tolerance in your own procedure decides the class, as the two examples above show. A related mistake is to check a balance only at one load. A step-by-step precision bench scale calibration covers linearity across the range, eccentricity and repeatability, and each test load needs a weight that passes the rule at its own nominal value.
Is NIST Class F still acceptable?
Class F tolerances were set in the 1990 edition of Handbook 105-1: 1 part in 10,000 for weights of 1 kg (2 lb) and larger, with wider proportional limits for small weights. A 50 lb Class F weight is allowed 2.3 g and a 25 lb weight 1.1 g. The NIST Handbook 105 series page lists the current edition.
For a quality manager, the practical points are these:
- Keep using calibrated Class F weights where they pass the one-third rule. Nothing in the 2019 handbook withdraws weights already in service.
- Buy to ASTM or OIML when you replace them. ASTM Class 6 tolerances match Class F at 100 g, 1 kg and 20 kg in the table above, but not at every size. NIST recommends against simply reclassifying Class F weights as an ASTM or OIML class.
- Write the edition into your procedure. “Class F per NIST Handbook 105-1 (1990)” is unambiguous. “Class F” alone now points at a table the current handbook does not contain.
How often should calibration weights be recalibrated?
NIST Good Measurement Practice 11 was written for laboratory standards, not for a customer’s check weights, but it is the most specific public guidance there is. It lets a laboratory lengthen an interval only on documented evidence such as calibration history and measurement assurance data. The same logic is in ILAC-G24 methods for setting calibration intervals.
What shortens the life of a weight is mostly handling:
- Bare hands on stainless steel laboratory weights. Skin oil adds mass and starts corrosion. Use gloves, forceps or a lifting fork.
- Sliding a weight across the pan, which wears both.
- Storing cast iron in a damp area. Rust and flaking paint change the mass in either direction.
- Magnetized tools or a magnetic stirrer near stainless weights.
Between calibrations, comparing two weights of the same nominal value on a good balance and trending the difference is a cheap intermediate check under ISO/IEC 17025 clause 6.4.10. It will not tell you which weight moved, but it tells you one of them did.
How does Techmaster calibrate scales, balances and weights?
Scales and balances are listed under Mass and Mass Related on Techmaster’s ANAB scope of accreditation (Certificate AC-1736) for Vista and Santa Clara, California, Orlando, Florida, and San Antonio, Texas. The published ranges and CMCs are on our mass and mechanical calibration scope page, and the service itself is described under ISO/IEC 17025 mass and mechanical calibration.
In our ten-year record of 381,916 calibrations, 31,077 are in the mass and mechanical discipline. Torque wrenches (4,269) and pressure gauges (2,378) lead that group, and weights account for 1,940 calibrations. The scale models we see most include the Ohaus SD75 and T71P, the Uline H-4593 floor scale and the Mettler Toledo PS60, all calibrated in our own laboratories.
Weights are handled differently, and we say so before quoting. Calibrating a mass standard is a separate measurement from calibrating a scale, and it is not a line on our own scope. In our records, nearly all customer weights were calibrated by a specialist mass laboratory under our subcontracting process. The certificate you receive identifies who performed the work. Our guide to subcontracted calibration and ISO/IEC 17025 accredited certificates explains what to look for on it.
When you send a scale or balance, tell us the tolerance you apply and the loads you check at. That lets us choose test loads that match your use and tell you before the work starts whether our accredited uncertainty at that load supports a conformity statement against your limit. A tolerance as tight as the two examples above can be close to a calibration laboratory’s own CMC, so check the scope for the range you need. It is the same information you need to choose your own check weights. See the full list of ISO/IEC 17025 accredited calibration services or review our ANAB accreditation and Certificate AC-1736.
Key takeaways
- For calibration weights, a test weight class is a tolerance on conventional mass plus construction rules. It is not a calibration.
- Choose the class from the device tolerance: weight error plus uncertainty under one-third of it, at the nominal value you use.
- ASTM and OIML classes do not map one to one. ASTM Class 1 sits between OIML E2 and F1.
- ASTM E617-23 changed Class 5 and 6 tolerances. Older charts are out of date.
- Using the certificate value instead of the nominal value lets a lower class meet a tighter need, if you control drift and data entry.
- NIST Class F is a legacy class. Keep calibrated weights in service where they fit, and replace them with ASTM or OIML classes.
Frequently asked questions
What class of weights do I need to calibrate an analytical balance?
For an analytical balance reading 0.1 mg, ASTM E617 Class 1 or OIML F1 is the common choice. Confirm it with the one-third rule: the weight’s tolerance at the load you check should be under one-third of the tolerance you apply to the balance. A tight in-house tolerance can require Class 0 or OIML E2.
Is ASTM Class 1 the same as OIML F1?
No. There is no official equivalence between the two systems. At 100 g, ASTM Class 1 allows 0.25 mg and OIML F1 allows 0.5 mg, so a Class 1 weight is tighter than F1 but looser than OIML E2 at 0.16 mg. Compare the tolerances at the nominal value you use.
Can I still use NIST Class F weights?
Yes, where their tolerance fits your application and they have a current calibration. Class F is a legacy class. The 2019 edition of NIST Handbook 105-1 no longer contains the Class F tables and directs new purchases for legal metrology to ASTM E617 or OIML R 111.
Is the class marking enough, or do my test weights need a calibration certificate?
The class marking is only a claimed tolerance. For traceability under ISO/IEC 17025 or ISO 9001, each weight needs a calibration certificate that shows its measured conventional mass, the measurement uncertainty and the date, from a laboratory competent to perform mass calibration.
How often should test weights be recalibrated?
No standard sets a single interval. NIST GMP 11 suggests initial intervals of 12 to 24 months for working mass standards, depending on the laboratory’s level, and 6 months for cast iron until stability data exist. Start there, then lengthen or shorten the interval from each weight’s calibration history and how it is handled.
Does Techmaster calibrate scales and balances under ISO/IEC 17025?
Yes. Scales and balances are on Techmaster’s ANAB scope, Certificate AC-1736, at our accredited laboratories in Vista and Santa Clara, California, Orlando, Florida, and San Antonio, Texas. Standalone test weights are calibrated through a specialist mass laboratory, and we tell you that before quoting.
Need a scale or balance calibrated against the tolerance you actually use?
Techmaster Electronics provides ISO/IEC 17025 accredited scale and balance calibration from laboratories in California, Florida and Texas. Toll free 1-866-779-5695.
Request a calibration quoteCalibration labs near you: Techmaster Electronics is an ISO/IEC 17025 accredited calibration laboratory (ANAB Certificate AC-1736) with accredited labs in Vista, CA, Santa Clara, CA, Orlando, FL and San Antonio, TX, plus a lab in Holly Springs, NC. Accredited parameters vary by lab; compare locations or request a quote.